Becton Dickinson (BD) is recalling all ethylene oxide (EtO) sterilised BD Venflon Pro Safety (VPS) and Venflon Pro IV Cannulae after identifying an increase in reports of leakage from the injection port.
-
BD have notified the MHRA that the sterility of some of their devices cannot be guaranteed due to quality issues with their third-party sterilisation provider.
-
Eaststone Limited has informed us of an issue related to underfilled or empty bottles for MidaBuc - Midazolam (as HCL) 10mg/mL Oromucosal Solution. The below batch is being recalled.
-
Gilead Sciences Ltd has informed us of a quality defect issue, in relation to specific lots of 5µm sterile filters, which are co-packed in cartons of the AmBisome product. The medical device is called Minisart Filter 16534-…
-
Guerbet Laboratories Limited is recalling certain batches of products as a precautionary measure due to reports received from healthcare professionals on the difficulties in attaching the Luer lock adapter to the Luer tip of…
-
Manufactured by Philips – defibrillator internal paddles may wear over time and might fail to deliver therapy, so it is important to do routine operational checks between each usage.
-
Manufactured by Abbott – cases of valvular insufficiency and early revision.
-
Manufactured by Philips – the rotary therapy selector switch may fail resulting in unexpected device behaviours which could lead to a delay or failure in delivering therapy.
-
The Philips HeartStart MRx Monitor/Defibrillator MRx may fail to identify a fault and alert the user in the event of internal damage suffered during a drop or due to severe mechanical shock.
-
All manufacturers – there have been reports of off-label modifications to haemofiltration systems when treating Covid-19 patients leading to serious injury and death.
-
Manufactured by Philips – due to a software or hardware issue the device may fail to start or deliver defibrillation therapy.
-
Manufactured by Ecolab/Microtek Medical Malta Ltd – bacterial contamination may cause an infection in patients.
-
GlaxoSmithKline, trading as Glaxo Welcome Ltd, is recalling all unexpired stock of Zantac (ranitidine hydrochloride) prescription only medicines (POM) from pharmacies.
-
Manufactured by Defiteq International BV or GGT Holding BV – the safety and performance of these devices cannot be verified and their safe use can no longer be assured.
-
Manufactured by Becton Dickinson (BD) – tubes may contain a hole or be damaged or deformed, potentially causing blood leakage and/or an inadequate blood-to-additive ratio leading to incorrect test results.
-
Manufactured by Philips – devices may lose power earlier than expected and users may not realise the loss of monitoring due to no alarm, which could contribute to a delay in emergency treatment.
-
Manufactured by Philips – this problem affects lithium ion batteries that have exceeded their specified replacement interval or number of charging cycles.
-
[Archived] Manufactured by Vyaire – Cartridges with an aluminium warming plate in the fluid pathway can lead to an IV infusion containing aluminium above currently recommended safe levels.
-
Aguettant Limited has informed the MHRA that there is an error in the Patient Information Leaflet (PIL) for batches of Atropine Sulfate 3mg/10ml Solution for injection in pre-filled syringe.
-
Manufactured by Medtronic Inc – a subset of dual chamber pacemakers may experience a loss of pacing therapy when programmed to a dual chamber mode with atrial-sensing (MDA/2019/008)
-
Actavis Group PTC EHF is recalling the below batches from pharmacies as a precautionary measure due to possible contamination with N‑nitrosodiethylamine (NDEA).
-
Manufactured by Eurotrol B.V – may give readings below the values assigned to the product, leading to incorrect measurement results.
-
Macleods Pharma UK Limited is recalling the below batches from pharmacies as a precautionary measure due to possible contamination with N‑nitrosodiethylamine (NDEA).
-
Manufactured by GE Healthcare – Device may stop ventilation when in PSV Pro Mode and users may be unable to change gas and agent settings when using End-Tidal Control.
-
Actavis Group PTC EHF is recalling the below batches from pharmacies as a precautionary measure due to possible contamination with N‑nitrosodiethylamine (NDEA).
-
List of field safety notices (FSNs) from medical device manufacturers from 17 - 21 December 2018
-
List of field safety notices (FSNs) from medical device manufacturers from 10 - 14 December 2018
-
Some valsartan containing products including certain batches supplied by Teva UK Limited and all unexpired batches supplied by Mylan are being recalled.
-
Philips M3538A lithium-ion batteries manufactured from 28 December 2017 to 20 March 2018 inclusive for the HeartStart MRx monitor/defibrillator may have an internal component failure.
-
IPS Specials are recalling specific batches of Melatonin 10 mg capsules because they have been assigned an incorrect extended expiry date.
-
Kyowa Kirin is advising caution in use for one lot of Bleo-Kyowa®, Powder for Solution for Injection, 15,000 IU following the detection of glass particles (EL (18) A/09)
-
FDC Pharma are recalling batch 087B087 due to an out of specification result for an unknown impurity during stability testing. (EL(18)A/08)
-
Caution in Use: Distribute to Pharmacy and Clinic Level.
-
Summary List of field safety notices (FSNs) from medical device manufacturers from 26 February – 02 March 2018
-
Caution in Use - Distribute to Hospital Pharmacy, Ward, Chemotherapy Unit and Clinic Level. (EL (18) A/05)
-
Glaxo Wellcome UK Limited is recalling one lot of Seretide Accuhaler 50/250mcg because a manufacturing issue has resulted in a small number of units not delivering the full number of doses. (EL (18) A/04)
-
Glaxo Wellcome UK is recalling two Lots of Ventolin Accuhaler 200mcg because a manufacturing issue with these Lots has been identified which has resulted in a small number of units not delivering the full number of doses (EL…
-
Manufactured by ICU Medical (formerly Hospira) – potential for connectivity engine module to disengage from main chassis, which may lead to power down without an alarm (MDA/2018/006).
-
The supply of batteries for these devices has been temporarily stopped so there may be a significant delay in the fulfillment of orders for new batteries.
-
Manufactured by Boston Scientific: specific models only – risk of syncope or pre-syncope due to periods of pacing inhibition from oversensing of the minute ventilation sensor signal (MDA/2018/001)
-
Caution in Use: Distribute to Hospital Pharmacy and Ward Level.
-
Caution in Use: Distribute to Pharmacy, Hospital, GP, Dental Practice, Ward and Clinic Level.
-
Summary List of field safety notices (FSNs) from medical device manufacturers from 27 November to 1 December 2017.
-
Caution in Use: Distribute to Hospital Pharmacy, Ward, Chemotherapy Unit and Clinic Level
-
Strathclyde Pharmaceuticals Ltd is recalling a single batch of Xarelto 20mg film-coated tablets (Rivaroxaban) because of reports of rogue blister strips of 15 mg tablets within two packs of 20 mg tablets.
-
(Focus Pharmaceuticals) Error with the barcodes on selected batches of 8 medicinal products – Class 4 Caution in Use (EL (17)A/08)
-
Specific manufacturers – measures to prevent inverted valve implantation resulting from incorrect mounting of the valve on its valve holder.
-
Torrent Pharma (UK) Ltd is recalling the below batches as a precaution because out of specification results for dissolution were obtained during routine stability testing.
-
Baxter Healthcare Limited are initiating a precautionary, company-led, ward-level recall of two batches of Sodium Chloride Infusion 0.9% w/v 1000 mL bags due to a potential leak with the bags.
-
Pharmacy and wholesaler level recall of two batches of Parallel Imported One-Alpha Drops x 10 ml because the dropper included in this pack delivers 0.05 micrograms of alfacalcidol instead of 0.1 micrograms as labelled on the…